Supplement for Infinity CentralStation Wide VG2.n and M300 VG1.n
Infinity M300
Modifications
ECG, arrhythmia, and ST segment monitoring
with M300
Monitoring paced patients
The amplitude to identify a pulse as a pacer pulse
is now:
Value
Amplitude
±2 to ±700 mV
SpO
and pulse rate monitoring with Masimo
2
and M300
SpO
precautions
2
Available settings refer to alarm grades. Available
grades are: Low (default), Medium, and High.
SpO
and pulse rate monitoring with Nellcor
2
and M300
SpO
precautions
2
Available settings refer to alarm grades. Available
grades are: Low (default), Medium, and High.
Technical specifications
Infinity M300
the folowing electrical technical specifications were
updated for the Infinity M300.
.
Value
Battery
Using Bedside Charger, approxi-
recharging
mately:
time
ECG only: 0 to 90 % = 11 to 13 hrs
ECG + continuous SpO
12 to 14 hrs
6
Battery life
Operating Concept
To transfer data from an M300 to another M300:
The following caution was added:
CAUTION
Do not select the same M300 as the source and
the destination for data transfer. This action will
result in discharging patient data on the M300.
M300 Offline Recordings
While M300 is offline or R50N recorder is not avail-
able, the last 5 recordings (automatic or manual)
are stored on M300. When the connection is re-
stored, those recordings will be sent to the R50N
recorder.
NOTE
The stored recordings on M300 are not sent to
ICS when the connection between M300 and ICS
is restored.
: 0 to 90 % =
2
Supplement Infinity CentralStation Wide VG2.n / M300 VG1.n
−
English
Value
With display and audio off, approxi-
mately:
ECG only: 22 hours
ECG + continuous SpO
: 18 hours
2