5. ERROR MESSAGE/MALFUNCTION
Follow the below recommended actions and press O/I button again to initiate
another measurement.
Error code
5.1 Other possible malfunctions:
The display shows the
symbol
The blood pressure
values displayed are
abnormally high or
low.
100042_MEDEL CHECK USER MANUAL_ ALL LANGUAGES.indd 29
Possible cause(s)
E 1
Deflation is too fast or the cuff
doesn't inflate
E 2
Deflation is too slow
E 3
The cuff pressure is excessive
E 20
The reading process cannot
detect the heart beat
E 21
Inaccurate reading
E Exx
There is a calibration problem
Problem
Re-apply the cuff and try again
Relax for a moment and try again
Movement can compromise the
measurement.
Relax for a moment and try
again .
Loosen the clothes on the arm and
try again
Relax for a moment and try again
Repeat the reading. If the problem
persists, contact the Service
Centre. Refer to the warranty for
the contact information and the
instructions concerning returns.
Possible cause(s)
Batteries are drained.
The cuff is not wrapped
correctly or not at the
level of the heart.
Too much stress is ap-
plied on your shoulder
or arm.
You move your arm
or arm muscles during
measurement.
Recommended Action
Recommended Action
Replace all batteries
with new ones.
Wrap the cuff correctly
and rise your hand so
that the cuff is at the
same level of your heart.
Relax yourself and make
measurement.
Remain still and do
not move /contract the
muscles during measure-
ment.