Medel CHECK Manual Del Usuario página 35

Ocultar thumbs Ver también para CHECK:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 108
The model Medel Check is intended for use in the electromagnetic environment
Immunity
test
Electrostatic
discharges
(ESD)
IEC 61000-4-2
Electrical fast
transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and
voltage
variations
on power
supply
input lines
IEC 61000-
4-11
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage before the application of the test level
100042_MEDEL CHECK USER MANUAL_ ALL LANGUAGES.indd 35
Guidance and manufacturer's declaration – electromagnetic immunity
specified below. The customer or the user of the model Medel Check should
ensure that it is used in such an environment.
IEC 60601 test level
± 6kV contact
± 8kV air
±2 kV for power
supply lines
.......
± 1kV in differential
mode
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT) for
5 sec.
3 A/m
Compliance level
Electromagnetic
environment - guidance
± 6kV contact
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
± 8kV air
material, the relative humidi-
ty should be at least 30 %.
Not Applicable
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Not Applicable
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Not Applicable
Mains power quality should
be that of a typical commer-
cial or hospital environment.
If the user of the model
Medel Check requires conti-
nued operation during power
mains interruptions, it is
recommended that the model
Medel Check be powered
from an uninterruptible
power supply or a battery.
3 A/m
Power frequency magnetic
fields should be at levels cha-
racteristic of a typical location
in a typical commercial or
hospital environment.
EN
35
16/05/2016 09:47:01

Publicidad

Tabla de contenido
loading

Tabla de contenido