Calibration Schedule; Glp Mode - BRAND Titrette Instrucciones De Manejo

Ocultar thumbs Ver también para Titrette:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

GLP Mode

The schedule for the next calibration can be stored in GLP Mode (GLP = Good Laboratory Practice).
What to do
1.
Bring up the
GLP mode
2.
Entering the
scheduled date
for calibration
3.
Confirming
setting
Note:
The stored scheduled date for calibration can be called up
any time the instrument is powered on. To do this, simply
press and hold the On/Off key. This brings up a continu-
ous display of GLP, and the year and month of the desired
scheduled date. Releasing the key ends the display, and
the instrument will be powered on. (If 'oFF' is selected
as the scheduled date for calibration, this deactivates the
function.)
How to do it
With the instrument powered
on, press and hold the
CLEAR key for more than
3 seconds. The following
modes will be repeatedly
shown in the display after
one another:
CAL - GLP - APO - dP.
When GLP appears in the
display, release the CLEAR
key. The input symbol blinks
and 'oFF' appears.
Press and hold the Pause
key until the desired date is
displayed. Pressing briefly
extends the scheduled
period stepwise. Pressing
the On/Off key shortens the
scheduled period. (Schedule
input can be from 'oFF' to
12/2099)
Press the CLEAR key in
order to confirm the input
of the scheduled date for
calibration.

Calibration Schedule

Keys to press
Display readout
> 3 s
+
1x
> 2 s
month
year
-
-
51

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido