Settings For Flow Measurements - STAUFF PPC-CAN-FR Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 123
4.4.2

Settings for flow measurements

For the flow measurements, the K-factor must be specified as well as the
measurement range. The K-factor is a proportionality factor which is de-
rived from the geometric shape of the flow sensor, the speed of flow and
the viscosity of the fluid. You can find the K-factor in the sensor's techni-
cal data sheet.
1 In the PC Configuration Program window, select the measurement
type Flow for flow measurement. The field Measurement Unit chang-
es to "L/min".
2 Enter the K-factor in the K-factor field which corresponds to your
measurement. The K-factor must not be zero (maximum value 99999).
The figure is accurate to 2 decimal places.
3 The minimum value (Range From field) is always zero and can not
be adjusted.
4 Select the appropriate maximum value for the sensor's range from the
drop-down menu under the Range To field. A list of these measuring
ranges can be seen in Section 4.1.
5 If you want to confirm the settings, press Save and then Exit to close
the program. If you do not wish to save the settings, press Exit.
50

Publicidad

Tabla de contenido
loading

Tabla de contenido