Fault detected
No ultrasonic vibration
No light
Leak between the
handpiece and the
handpiece cord
Leak between the
handpiece cord and the
device connector
Liquid leaking under the
device
Liquid leaking from the
tank connector
Air leak
X - ELECTROMAGNETIC COMPATIBILITY
Important:
The power cord, the handpiece cord and the footswitch cord must be kept apart.
The P M
N
XS requires special precautions to be taken with regard to electromagnetic
AX
EWTRON
compatibility.
It must be installed and prepared for use as described in chapter 4.
Certain types of mobile telecommunication devices such as mobile telephones are likely to interfere
with the P M
N
AX
EWTRON
The recommended separation distances in this section must therefore be complied with.
The P M
N
XS must not be used near or on top of another device.
AX
EWTRON
If this cannot be avoided, its operation under the conditions of use must be checked beforehand.
Possible causes
Incorrect tip tightening.
Faulty connector contact.
Handpiece cord wire cut.
Handpiece light ring is missing.
Faulty light ring.
Faulty light ring connector contacts.
Light ring reversed polarity.
Faulty cord / handpiece connector
contacts.
Other.
Worn handpiece 1.15x1 seal.
Worn hydraulic connections 1.15x1 seal. Replace the seal.
Pipe rupture.
Worn seals.
Worn seal.
XS.
Solutions
Retighten the tip with the wrench.
Clean the connector contacts.
Replace the cord.
Install the light ring.
Replace the light ring.
Clean the connector light ring contacts.
Position the light ring by lining up the lug
Clean the handpiece and/or connector
contacts.
Contact your installer-integrator.
Replace the seal (kit F12304).
Return to SATELEC after-sales.
Replace the seals.
Replace the seal.