Electromagnetic Emissions; Electromagnetic Immunity - Satelec Acteon P MAX Newtron XS Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 40
The use of accessories other than those specified or sold by SATELEC as replacement parts may have
the consequence of increasing the emissions or decreasing the immunity of the P M

10. 1 ELECTROMAGNETIC EMISSIONS

The P M
N
XS is intended for use in the electromagnetic environment specified in the table
AX
EWTRON
below.
The user and/or installer must ensure that the P M
Emission test
RF emission - CISPR 11.
RF emission - CISPR 11.
Harmonic current emission
IEC 61000-3-2.
Voltage fluctuation and flicker
IEC 61000-3-3.

10. 2 ELECTROMAGNETIC IMMUNITY

The P M
N
XS is intended for use in the electromagnetic environment specified in the table
AX
EWTRON
below.
The user and/or installer must ensure that the device is used in such an electromagnetic
environment.
Immunity test
Electrostatic
± 6KV contact
discharge (ESD)
± 8KV air
IEC 61000-4-2.
Electrical fast
transients
± 2KV for power supply lines
IEC 61000-4-4.
Surges
± 1KV differential mode
IEC 61000-4-5.
± 2KV common mode
<5% UT (>95% dip in UT) for 0.5
cycles.
40% UT (60% dip in UT) for 5
Voltage dips, short
cycles
interruptions and
voltage variations
70% UT (30% dip in UT) for 25
IEC 61000-4-11.
cycles
<5% UT (>95% dip in UT) for 250
cycles
18
Compliance
Group 1
Class B
Class A
Complies
IEC 60601 test level
N
AX
EWTRON
Electromagnetic environment - guidance
P M
N
XS uses RF energy for internal operation.
AX
EWTRON
Therefore, its radiofrequency emissions are very low and are not likely
to cause any interference in nearby equipment.
P M
N
XS is suitable for use in all establishments, including
AX
EWTRON
domestic establishments and those directly connected to the public
low voltage power supply network that supplies buildings used for
domestic purposes.
Compliance level
± 6KV contact
± 8KV air
± 2KV for power supply lines
± 1KV differential mode
± 2KV common mode
<5% UT (>95% dip in UT) for 0.5
cycles.
40% UT (60% dip in UT) for 5
cycles
70% UT (30% dip in UT) for 25
cycles
<5% UT (>95% dip in UT) for 250
cycles
XS is used in such an environment.
Electromagnetic environment - guidance
Floors must be wood, concrete, cement or tiled.
If floors are covered with synthetic material
(carpet, etc.), the relative humidity must be at
least 30%.
Power quality should be that of a typical
commercial or hospital environment (hospital,
clinic).
Power quality should be that of a typical
commercial or hospital environment.
Power quality should be that of a typical
commercial or hospital environment.
If the use of the P M
N
AX
continued operation during a power cut, it is
recommended that the product be powered from a
separate power supply (UPS, etc.).
N
XS.
AX
EWTRON
XS requires
EWTRON

Publicidad

Tabla de contenido
loading

Tabla de contenido