Installation/First Use; Unpacking The Device; Recommendations; Installation - Satelec Acteon P MAX Newtron XS Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 40
Overall dimensions:
Control unit:
Width: 270mm
Height: 100mm
Height with tanks: 131mm
Depth: 246mm
Weight: 3Kg
Footswitch:
Width: 173mm
Height: 46mm
Depth: 177mm
Cord length: 2,000mm
Weight: 610g
Handpiece cord: 2,000mm
Tank capacity: 300ml each
Temperatures:
Operating: +10°C to +40°C.
Storage: -20°C to +70°C.
Humidity:
Operating: 30% to 75%.
Storage: 10% to 100% including condensation.
IV - INSTALLATION / FIRST
USE

4. 1 UNPACKING THE DEVICE

Upon reception of the device, check for any
damage caused in transit.
Contact your supplier if necessary.

4. 2 RECOMMENDATIONS

Have your connections to the electrical and air
supplies made by a certified dental installation
technician.
The electrical connection of the P M
must comply with applicable standards. A
grounded power supply socket must be used.
The device must be connected to the dental
office's medical-quality air distribution supply.

4. 3 INSTALLATION

Important:
Never rotate the handpiece connector on its
cord or you may damage the device.
Do not place the P M
on top of another device.
Do not intertwine the power cord with the
footswitch cord or the handpiece cord.
- Place the control unit in position on a fixed,
horizontal surface or with a slope of no more
than 5°.
- Check that the power switch is in position O
(off).
- Connect the power cord to the receptacle of
the device.
- Connect the power cord to a socket with a
ground pin.
- Connect the multifunction footswitch to the
footswitch connector. Fully tighten the screws.
- Position the footswitch so that it is easily
accessible for your feet.
- Check that there is no humidity in the
handpiece cord connector connections. If there
is, dry it out by wiping, then blowing with the
multifunction syringe.
- Connect the handpiece cord to the connector
on the device located on the front panel,
making sure that the two marks on the cord
plug and the connector symbol
N
AX
EWTRON
N
XS close to or
AX
EWTRON
are lined up.
XS
9

Publicidad

Tabla de contenido
loading

Tabla de contenido