OBJ_DOKU-189631-008.fm Seite 20 Donnerstag, 11. März 2021 9:58 09
5.2
Permitted settings on the Erbe VIO
For a monopolar CUT mode:
• Art. No. 20150-060, 20150-061:
– Effect: Effect with an HF peak voltage (U HF)
1)
2,500 V
p
– Power: maximum 80 watts
• Art. No. 20150-062:
– Effect: Effect with an HF peak voltage (U HF)
1)
770 V
p
– Power: maximum 80 watts
For a monopolar COAG mode:
• Art. No. 20150-060, 20150-061:
– Effect: Effect with an HF peak voltage (U HF)
1)
2,500 V
p
– Power: maximum 80 watts
• Art. No. 20150-062:
– Effect: Effect with an HF peak voltage (U HF)
1)
990 V
p
– Power: maximum 80 watts
1.) The HF peak voltage of an effect depends on the mode selected and
the power setting (watts) used. More information about selecting an
appropriate effect is available in the user manual of the VIO HF elec-
trosurgical unit.
20