Hold the Chip by the yellow part so that the
"microINR" inscription can be read correctly.
Insert the Chip into the slot and push it until
it stops.
• Make sure the Chip has reached the end.
If the pouch of the Chip is open, damaged or the film
of the Chip is removed, please, dispose of the
Chip and use a new one.

Modes of Use

The microINR system provides two application modes of the
sample:
- Approaching the
sample to the Meter.
iLine Microsystems only recommends the option of approaching
the Meter to the sample for users who are experienced in
conducting INR assays with the microINR system.
- Approaching the Meter to
the sample.

Conducting the Test

• Once the Chip is inserted, the
Meter will perform the quality
controls mentioned in section 2.6
of these instructions for use. After
conducting the aforementioned
quality controls prior to the test,
the "control" symbol will light up. If
the first level of quality control is not passed, an error
message will be displayed on the screen of the Meter.
• The Chip begins to flash and heats
up until it reaches the proper
temperature. Once this temperature is
reached:
- The device emits an audible
signal (beep).
- The drop symbol begins to
flash on the display.
- A count down appears (80s).
- The Chip emits a steady light.
• Perform the fingerstick (see section 3.3 of these instructions
for use).
• Make sure to obtain a spherical and properly sized drop,
equivalent to a teardrop.
• Apply the drop of blood on the Chip immediately, by contact
with the entry channel, without resting the finger on the Chip.
• The Meter will emit a beep when it detects the sample for
the test and the drop symbol will stop flashing.
09