Instructions for Use
Set-up
The TwisTorr 305-ICQ & 305-IC SF is an electronic device sensitive to
CAUTION
electrostatic discharges, and, to this point, it is supplied in a protective bag with
caps on the USB connection as well as on the DB15 connection. Remove these
protective caps only when connecting the pump to the controller.
Do not remove the adhesive and protective cap until you are ready to install the
CAUTION
turbopump to the system.
Figure 32
Install this device indoors only, and do not use the pump in an environment
exposed to atmospheric agents (rain, ice, snow), dust, aggressive gases, or in
explosive environments or those with a high fire risk.
During operation, the following environmental conditions must be complied with in
order to avoid condensation:
maximum ambient pressure: 2 bar above atmospheric pressure
temperature: from +5 °C to +35 °C
relative humidity: 0 – 90 % (non-condensing).
188
Agilent TwisTorr 305-ICQ & 305-IC SF User Manual