Warning Messages - Agilent TwisTorr 305-ICQ Manual Del Usuario

Tabla de contenido

Publicidad

Technical Information

Warning messages

Table 10 Warning messages
Message
Description
CK_RTC_RESET
Real Time Clock malfunction
A1 THERMALFAIL
A1 Accessory Output shut
down due to its
overtemperature protection
intervention
B1 THERMALFAIL
B1 Accessory Output shut
down due to its
overtemperature protection
intervention
DO1 THERMALFAIL
DO1 Digital Output shut down
due to its overtemperature
protection intervention
DO2 THERMALFAIL
DO2 Digital Output shut down
due to its overtemperature
protection intervention
V24VDC WARNING
+24VDC OUTPUT (V+), too
UNDERVOLTAGE
high load condition applied to
the output
ANALOGOUT OUT
AO1 Analog Output, too high
TOLL
load condition applied to the
output
Agilent TwisTorr 305-ICQ & 305-IC SF User Manual
Action
Contact Agilent to get the
appropriate directions.
Power off the unit in order to allow
it to cool down quickly.
Remove all the loads which are
applied to the output; check the
loads electrical ratings (they have
match with the output
specifications); check the loads for
short circuits absence/lack of
insulation.
Power on the unit, without any
load, the yellow led light must be
turned off.
If YES, reconnect the load.
If NO, contact Agilent to get the
appropriate directions.
Power off the unit.
Remove all the loads which are
applied to the output; check for the
loads electrical ratings (they have
to match with the output
specifications); check for the loads
absnce of short circuits/lack of
insulation.
Power on the unit, without any
load connected to the output; the
yellow led light must be turned off
immediately.
If YES, reconnect the load.
If NO, contact Agilent to get the
appropriate directions.
Remove all the loads which are
applied to the output; the yellow
led light must be turned off
immediately.
If YES, before reconnecting the
load to the output, check it for its
electrical ratings (they have to
match with the output specifica-
tions) and check it for the absence
of short circuits/lack of insulation.
If NO, contact Agilent to get the
appropriate directions
Led Status (for ref)
Red
Yellow
Green
OFF/ON
ON, Solid Blinking/
Solid
Flashing/
Inv.Flashing/
Solid
OFF/ON
ON, Solid Blinking/
Solid
Flashing/
Inv.Flashing/
Solid