Instructions for Use
Installation of ISO-K Flange
For ISO-K flange connections, fix the two flanges with the clamps model
IC63250DCMZ or equivalent, as shown in the following figure (a protective screen
can optionally be used).
Figure 33
Use the required number of claw clamps:
4 for ISO 100 flange
4 for ISO 160 flange
Tighten the bolts with a torque of 22 Nm.