Agilent TwisTorr 305-ICQ Manual Del Usuario página 249

Tabla de contenido

Publicidad

Technical Information
For example:
reference quantity: frequency
threshold: 500 Hz
hysteresis: 1%
activation type: "high level"
Figure 85
The set point output stays at 0 Vdc until the frequency becomes higher than 505 Hz
(that is 500 Hz + 1% of 500 Hz), then the output goes at 24 Vdc and stays at 24 Vdc
until the frequency be-comes lower than 495 Hz (that is 500 Hz – 1% of 500 Hz).
It is possible to delay the set point checking for a programmable delay time.
The PROGRAMMABLE SET POINT signal has the following default settings:
reference quantity: frequency
threshold: 909 Hz
hysteresis: 2 %
activation type: high level
delay time: 0 second.
The TwisTorr 305-ICQ & 305-IC SF Software (optional) allows the operator to set
NOTE
all the programmable feature.
Agilent TwisTorr 305-ICQ & 305-IC SF User Manual
Programmable setpoint
249

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Twistorr 305-ic sf

Tabla de contenido