Approvals; Marking; Docking Station; Approvals Issued - Dräger REGARD 3000 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

|
en
System overview
3.6

Approvals

REGARD 3000 is approved under IEC/EN 61508:2010 and EN 50402:2017 for
SIL 2 applications or for SIL 3 with the corresponding system layout. For details
about the configuration and system layout for SIL 2 or SIL 3, see the Dräger
REGARD 3000 Safety Manual.
REGARD 3000 is approved for measuring
– combustible gases and vapours in line with ATEX
– oxygen inertisation measurements in line with ATEX
– oxygen and toxic gases
3.6.1

Marking

REGARD 3000

Docking Station

BVS 21 ATEX G 001 X
PFG 21 G 002 X
www.draeger.com
Dräger Safety AG & Co. KGaA
23560 Lübeck, Germany
3.6.2

Approvals issued

Measurement technology for combustible gases and oxygen (inertisation
measurement)
Test body:
Standards:
Test number:
104
Docking station
Sach-Nr. /
3705683
Order code
Fabr-Nr. /
Serial no.
0158
II (1) G
10
II (1) D
QR-
code
Made in Germany
DEKRA Testing and Certification GmbH
EN 60079-29-1:2016
EN 50104:2019
EN 50271:2018
BVS 21 ATEX G 001 X
Base Unit
REGARD 3000
apply label here/ Etikett hier aufkleben
apply label here/ Etikett hier aufkleben
apply label here/ Etikett hier aufkleben
Input 100-120 V / 200-240V~, max 1A
Output
65W
Frequency 50 - 60 Hz
BVS 21 ATEX G 001 X
PFG 21 G 002 X
II (1) G
II (1) D
IP65
0158
Dräger Safety AG & Co. KGaA, 23560 Lübeck, Germany
www.draeger.com
|
Instructions for use
REGARD 3000
10
Made in Germany