Table 2
Guidance and manufacturer's declaration – electro-
magnetic Immunity
Immunity test
IEC 60601-1-2
Test level
Electrostatic
±8 kV contact
discharge (ESD)
±2 kV, ±4 kV,
IEC 61000-4-2
±8 kV, ±15 kV air
Electrical fast
transient/burst
Not applicable
IEC 61000-4-4
Surge
Not applicable
IEC61000-4-5
Voltage dips,
short interrup-
tions and volta-
ge variations on
Not applicable
power supply
input lines
IEC 61000-4-11
Power frequen-
cy magnetic
30 A/m
field
50Hz/60Hz
IEC 61000-4-8
Compliance
level
±8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15 kV air
Not applicable
Not applicable
Not applicable
30 A/m
50Hz/60Hz
233
Conduced RF
Not applicable
IEC61000-4-6
Radiated RF
10 V/m
IEC61000-4-3
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE U
is the a.c. mains voltage prior to application of
T
the test level.
Not applicable
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz