Final test values related to the nominal capacity (maximum volume) indi-
cated on the instrument and the indicated volume steps, obtained when
instrument and distilled water are equilibrated at ambient temperature
(20 °C/68 °F) and with smooth operation. According to DIN EN ISO 8655.
Accuracy tolerances for the Transferpette
Volume range
µl
0.5 -
10
2 -
20
10 - 200
50 - 1000
250 - 5000
* A = Accuracy, CV = Coefficient of Variation
Note:
The device is marked in accordance with the German Measurement and
Calibration Act as well as the Measurement and Calibration Regulation:
19
Character string: DE-M (DE for Germany), framed by a rectangle, as well as the
last two digits of the year in which the marking was affixed (here: 2019).
Note:
Testing instructions (SOPs) and a demo version of the EASYCAL
software are available for download at www.brand.de.
®
Volume step
A*
≤ ± %
µl
10
1.0
5
1.5
1
5.0
20
1.0
10
1.5
2
5.0
200
0.8
100
1.2
20
4.0
1000
0.6
500
1.0
100
3.0
5000
0.6
2500
1.0
500
3.0
electronic
CV*
Increment
≤ %
µl
0.4
0.01
0.8
2.0
0.4
0.02
0.8
2.5
0.2
0.2
0.3
0.6
0.2
1.0
0.3
0.6
0.2
5.0
0.3
0.6
™
4.0 calibration
Accuracy Table
20 °C
!
Ex
Recommended
type of tip, µl
0.5 -
20
0.5 -
20
2 -
200
50 - 1000
500 - 5000
57