Descargar Imprimir esta página

Stryker 6254 Manual De Uso página 28

Ocultar thumbs Ver también para 6254:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 123
Phenolic cleaners (active ingredient - o-phenylphenol)
Chlorinated bleach solution (5.25% - less than 1 part bleach to 100 parts water)
Avoid oversaturation. Do not allow the product to stay wet longer than the chemical manufacturer's guidelines for proper
disinfecting.
N N o o t t e e
• Failure to follow the above directions when using these types of cleaners may void this product's warranty.
• Always wipe the product with clean water and dry after cleaning. Some cleaning products are corrosive in nature and
may cause damage to the product. Failure to properly rinse and dry the product leaves a corrosive residue on the
surface of the product and may cause premature corrosion of critical components.
P P r r e e v v e e n n t t i i v v e e m m a a i i n n t t e e n n a a n n c c e e
Remove product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks
more often based on your level of product usage. Service only by qualified personnel.
N N o o t t e e - - Clean and disinfect the exterior of the support surface before inspection, if applicable.
O O p p e e r r a a t t i i o o n n
Cleaning and disinfecting
Inspection
S S t t a a i i r r - - T T R R E E A A D D system belt
reconditioning
Inspect the following items:
All fasteners secure (reference all assembly drawings)
All welds intact, not cracked or broken
No bent or broken tubing or sheet metal
No debris in wheels
All wheels secure and rolling
Front casters secure, rolling, and swiveling (if equipped)
Wheel locks hold wheels securely when on and clear the wheels when off
Chair unfolds and locks properly
No rips or cracks in seat or backrest
Restraint straps intact and working
Foot end carrying handles extend and lock in place
Head end carrying handles fold and unfold
Upper control handle extends and locks in all positions
S S t t a a i i r r - - T T R R E E A A D D system mechanism unfolds and locks in place (if equipped)
S S t t a a i i r r - - T T R R E E A A D D system belt rolls (if equipped)
S S t t a a i i r r - - T T R R E E A A D D system belt inner cords not showing; replace if necessary (if equipped)
S S t t a a i i r r - - T T R R E E A A D D system performs as desired; recondition belts if necessary (if equipped)
No lubricants present on the S S t t a a i i r r - - T T R R E E A A D D system belts or the track frame surfaces
EN
24
S S c c h h e e d d u u l l e e
Each use
Inspect chair every 6 months
After usage on approximately 500
flights of stairs or if the S S t t a a i i r r - - T T R R E E A A D D
system performance is sluggish
P P r r o o c c e e d d u u r r e e
See Cleaning in the Operations
Manual
See below for checklist
See Track belt reconditioning in the
Maintenance Manual
6254-109-005 Rev C.0

Publicidad

loading