Reprocessing
The camera console is not intended to come into contact with the patient.
It may be cleaned, but not sterilized. The camera head and coupler may
contact the patient and should both be cleaned and sterilized prior to every
use. (Refer to the appropriate coupler user guide—identified in the "Product
Description" section of this user guide—for instructions on reprocessing the
coupler.)
Cleaning the Console
Should the camera console need cleaning, wipe it down with a sterile cloth
and mild cleaning solution.
Disconnect the console from the AC power source before
cleaning.
Never immerse or sterilize the camera console as this will
damage the camera and void the warranty.
Cleaning, Disinfecting, and Sterilizing the Camera Head
These reprocessing instructions are provided in accordance with ISO 17664,
AAMI TIR12, AAMI ST79, and AAMI ST81. While they have been validated by
Stryker as being capable of preparing the device for re-use, it remains the
responsibility of the processor to ensure that the reprocessing as actually
performed, using equipment, materials, and personnel in the reprocessing
facility, achieves the desired result. This normally requires validation and
routine monitoring of the process. Stryker recommends users observe these
standards when reprocessing medical devices.
Warnings
• This device must be cleaned and sterilized prior to the first use and
after every subsequent use.
• Use only the sterilization cycles outlined in this document. Using
unspecified sterilization cycles may damage the device or result in
incomplete sterilization.
• Separate the camera head, coupler (1488210105 and 1488710105
only), and scope prior to cleaning, disinfection, or sterilization. If the
coupler and camera head are cleaned, disinfected, or sterilized as a
single unit, disconnecting the coupler during use will compromise
the sterility of the two products. (Refer to the coupler and scope
product manuals for reprocessing instructions.)
• Wear appropriate protective equipment: gloves, eye protection, etc.
• Sterilize only one camera head per tray, or incomplete sterilization
may result. Follow any instructions provided with the sterilization
tray or system regarding tray setup and other devices that may be
sterilized within the same tray.