Descargar Imprimir esta página

User Maintenance; Using Sterile Drapes; Replacing The Fuses; Periodic Maintenance Schedule - Stryker 1488 Serie Manual Del Usuario

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 249

User Maintenance

Using Sterile Drapes

Using sterile drapes will ensure maximum longevity of your 1488 HD
Camera Head. For best results, follow the instructions provided by the drape
manufacturer.

Replacing the Fuses

To avoid the risk of fire, use only fuses of the value specified on
the fuse label located on the rear panel of the console.
1.
Unplug the power cord from the wall outlet and remove the cord from
the transmitter console.
2.
Unlatch the fuse holder above the AC inlet and remove it. (You may need
to press the tab on the fuse holder with a slender screwdriver to release
the latch.)
3.
Replace the fuse with the same value and rating.
4.
Reinstall the fuse holder until the tab snaps in place.

Periodic Maintenance Schedule

To ensure safe operation of the Model 1488 HD Video Camera
you should periodically perform the following procedure:
Every 12 months, check the earth leakage current to <500 µA (<300 µA
in USA), ground protective earth impedance to <0.1 ohms, and power
consumption less than or equal to rated power. Use a true RMS digital
multimeter and safety analyzer to perform this test.
Note: Refer calibration and operating difficulties not detailed in this manual
to your Stryker Endoscopy sales representative.

Disposal

This product contains electrical waste or electronic equipment.
It must not be disposed of as unsorted municipal waste and
must be collected separately in accordance with applicable
national or institutional related policies relating to obsolete
electronic equipment.
Dispose the 1488 HD according to local laws and hospital practices.
EN-36
All manuals and user guides at all-guides.com

Publicidad

loading

Este manual también es adecuado para:

14880100001488010001148821010514886101221488710105