Guidance and Manufacturer's Declaration: Electromagnetic Immunity
1488 HD Camera is intended for use in the electromagnetic environment specified below.
The customer or the user of 1488 HD Camera should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Conducted RF
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
80MHz to 2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
Compliance
Level
3 Vrms
3 V
3 V/m
3 V/m
Electromagnetic Environment:
Guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the 1488 HD Camera
system, including its cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended Separation
Distance
d = 1.17 √P
d = 1.17 √P
80 MHz to 800 MHz
d = 2.23 √P
800 MHz to 2.5
GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey (a) , should be less than
the compliance level in each
frequency range (b) . Interference
may occur in the vicinity of
equipment marked with the
following: