4.4
Conducting a performance test
Before every use conduct a performance test.
4.5
Footswitch pedals and their function
Footswitch 20188-350:
• Blue pedal: activates coagulation
• Black ReMode button: activates the switch between up to
6 setting options in a program
Footswitch 20189-350, 20189-351, 20189-352, 20189-353:
• Yellow pedal: activates cutting
• Blue pedal: activates coagulation
• Black ReMode button: activates the switch between up to
6 setting options in a program
5
Cleaning and disinfection
5.1
Safety instructions
Do not exceed a temperature of max. 75 °C!
Erbe recommends wiping or spray disinfection with an appropri-
ate disinfectant. Comply with the disinfectant manufacturer's
stipulations as to its suitability and concentration.
Do not sterilize!
5.2

Manual cleaning and disinfection

Rinse the product with water immediately after use and wash it
with an alkaline cleaning solution.
Disinfect the product after cleaning. When doing so, always check
the instructions provided by the disinfectant manufacturer to en-
sure that the agent is suitable and of the correct concentration.