Metrohm 798 MPT Titrino Instrucciones De Uso página 105

Ocultar thumbs Ver también para 798 MPT Titrino:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

798 MPT Titrino
OFF means that the criterion is not monitored.
3.2.2.22. Mode.Parameter.StopCond.FillRate
Parameters for DET, MET, SET: Filling rate in the titration in mL/min. Max.
means maximum possible filling rate with the Exchange Unit in current use.
3.2.2.23. Mode.Parameter.Statistics.Status
Mode.Parameter.Statistics.MeanN
Mode.Parameter.Statistics.ResTab.Selected
Mode.Parameter.Statistics.ResTab.DelN
Entries for the statistics calculations.
.Status:
On/off switching. Requirement for statistics calculations is a valid
assignment, see 3.2.2.9.
.MeanN:
Number of individual results for statistics calculations.
.ResTab.Select: Selection of the table for the statistics calculations.
original: Original table. The original table is (again) set up, i.e. any
individual results which have been deleted are reincorporated
in the statistics calculations.
delete n: Single result lines are removed from the statistics calcu-
lation. All results of the corresponding line in the statistics
table are deleted. Specification of the line number in
.ResTab.DelN.
delete all: Clear entire statistics table. The results can not be reac-
tivated.
.ResTab.DelN: Specification of the line number to be deleted.
3.2.2.24. Mode.Parameter.Evaluation.EPC
Mode.Parameter.Evaluation.Recognition.Selected
Mode.Parameter.Evaluation.Recognition.Window.1.LowLim
Mode.Parameter.Evaluation.Recognition.Window.1.UpLim
etc. up to 9 windows
Mode.Parameter.Evaluation.FixEP.1.Value pH:
etc. up to 9 fix EP's
Mode.Parameter.Evaluation.pK
Parameters for DET and MET: Evaluation of the EP's, see page 20.
.EPC: EP criterion in pH, in mV (with U and Ipol) or in µA (with Upol).
.Recognition.Selected: EP recognition.
all: All endpoints found are recognized.
great: Only the largest EP is recognized.
last: Only the last EP is recognized.
3.2 Remote control commands
0.01...150, max.
original, delete n,
DET: 0...5...200
MET pH: 0.1...0.50...9.99
U, Ipol:
Upol:
0.1...2...99.9
all, greatest,
last, window, OFF
pH:
0...±20.00, OFF
U, Ipol:
0...±2000, OFF
Upol:
0...±200.0, OFF
Input range as LowLim
0...±20.00, OFF
U; Ipol: 0...±2000, OFF
Upol: 0...±200.0, OFF
ON, OFF
2...20
delete all
1...20
1...30...999
ON, OFF
101

Publicidad

Tabla de contenido
loading

Tabla de contenido