Metrohm 798 MPT Titrino Instrucciones De Uso página 128

Ocultar thumbs Ver también para 798 MPT Titrino:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

3.2. Remote control commands
124
.Grid:
On/off switching of grid over curve.
.Frame:
On/off switching of frame surrounding the curve. If grid and frame
are switched off, the curve is printed faster as the print head does
not have to move to the end of the paper.
.Scale:
Type of scaling of the measured value axis: Full means that the
scale runs from the smallest up to the greatest measured point.
With auto, the smallest measured value is taken and the next
smaller tick defines the beginning of the scale; the next greater
tick to the greatest measured value is the end of the scale.
.Right:
Relative specification of the width of the output medium (e.g. pa-
per width) for the length of the measured value axis. 1 means the
measured value axis is plotted over the entire width of the paper
(largest possible width). In extreme cases, the writing of the right
tick may lie outside.
.Feed:
Length of the volume axis referred to the buret cylinder volume,
V(B) per cm (0.1 means, e.g. 1 mL/cm with a 10 mL Exchange
Unit). Depending on the printer, the measure in cm may not al-
ways be correct.
3.2.2.127. Setup.PowerOn
Simulation of 'power on'. The device has the same status as after power on:
The cylinder is filled, error messages deleted and the current sample number
set to 0. The method last used is ready for operation.
3.2.2.128. Setup.Initialise
Setup.Initialise.Select
Setting of default values for the following areas:
ActMeth:
Current method. Parameters, calculations, and assignments for
the data output, operands C01...C19.
Silo:
The silo memory is deleted. Same function as delete entire silo.
Calib:
pH calibration data for all measuring inputs.
Config:
All values under &Config.
Assembly: All values under &Assembly.
Setup:
All values under &Setup.
All:
Values of the entire tree (except silo and method memory).
The action must be triggered with &Setup.Initalise $G.
3.2.2.129. Setup.RamInit
Initializes instrument, see page 148. All parameters are set to their default
value and error messages are cleared. The user and silo memories will be de-
leted. The user memory contains the default user methods from Metrohm.
3.2.2.130. Setup.InstrNo
Setup.InstrNo.Value
Instrument identification for report output.
Set the value with &Setup.InstrNo $G .
ActMeth, Silo, Calib, Config,
Assembly, Setup, All
serial number, 8 ASCII characters
$G
$G
$G
$G
798 MPT Titrino

Publicidad

Tabla de contenido
loading

Tabla de contenido