2 General safety instructions
The HF device may be used only by trained medical staff. The
surgeon and medical staff must be trained in the fundamental
principles, rules for use and risks of HF surgery and must be
familiar with them.
Read the operating manual carefully and thoroughly before
using the device!
Use aseptic techniques when removing the product from
the packaging.
Do not use the product if the packaging is opened or
damaged, the product will no longer be sterile. Do not re-
sterilize!
The product is intended for single use and may not be
resterilized. While autoclaving, the instrument gets
thermally damaged, deformed so that it loses its function.
Do not use the product if its expiry date has elapsed.
Do not use defective instruments.
2.1 HF device
Use only approved HF devices and programs (see section
8, page 13).
Observe the operating instructions of the HF device and
the general instructions for electrosurgical operations!
Improper use of HF current can lead to endogenous and
exogenous burns and explosions:
Avoid direct skin contact with HF cables.
Avoid contact with flammable gases and liquids.