2.2 HF cable
Improper use of HF cables can lead to patient injuries:
NОvОr ХКв ЭСО HŻ МКЛХО on ЭСО pКЭТОnЭ's skТn.
When plugging or unplugging the HF cable, always grasp
the connector directly.
Use only HF cables that are in perfect condition. Never
use a defective HF cable.
The HF cable may cause interference to imagery on monitors:
Never route the HF cable alongside a camera cable.
Do not lay the HF cable in loops.
Consult the operating manuals of the BOWA HF
generators for additional information on interference with
other devices.
2.3 Active electrodes and instrument
Defective or worn electrodes can cause burns on patients:
Never use or repair worn or defective electrode surfaces.
Dispose of them.
Hot electrode surfaces may cause patient injuries:
Maintain sufficient distance between the tips of the
instrument and sensitive tissue structures, such as the
pancreas or intestine.
Ensure that hot instruments are not used for preparation.
Inadvertent activation of the ligation instrument may cause
patient injuries:
Do not lay the ligation instrument on the patient.
Dirty electrodes may cause a short circuit, thereby resulting in
functional failure of the ligation instrument.
Clean the jaw electrodes regularly with a moist cloth.