4 Operation
WARNING
Risk of patient injury!
Risk of patient injuries from the combustion or
explosion of flammable liquids and gases!
4.1 Before use
1.
Switch on the HF device.
2.
Remove the product from the sterile packaging.
3.
Connect the HF cable to a bipolar socket of the HF device
4.
The output power and mode is automatically set on the
BOWA ARC generator with ЭСО PХuР'n CuЭ COMŻORT
instrument detection.
5.
Perform a thorough visual inspection and functional test
before using the ligation instrument.
4.2 Inspection
1.
Check if the rotation wheel can be turned easily until the
stop is reached.
2.
Check the jaws for proper opening and closing using the
lever.
3.
Lock the handle by pressing the lever down.
4.
Pull the trigger to check for ease of motion.
5.
Check the finger switch for proper activation.
The activation signal sounds while the finger switch is
being pushed.
MN031-600-S0 EN
Use only intact products.
Avoid contact with flammable gases and liquids, such
as skin cleaners, disinfectants and anaesthetic gases
ERGO 310D
9