2. Next, the sleeve
measurement has been reached. Alternatively, you can create this pressure
manually by pressing and holding the START button
has automatically reached a pressure of at least 30 mmHg. When you think
that the device has reached the ideal starting pressure for a measurement,
release the START button
numerically.
3. The measuring process begins automatically when inflation is complete by
releasing the pressure built up in the cuff. The display shows the current
pressure in the cuff.
NOTE
If you press the START button
the inflation process stops. The device switches off and the air
escapes from the sleeve.
4. Initially, deflation occurs at a constant speed. The pulse symbol
appears on the display. As soon as the pulse can be determined, deflation
is synchronised with the beating of the heart until the measurement ends.
The device beeps for each heartbeat during this procedure. At the same
time, the pulse symbol
5. Once the measurement has been taken, the sleeve suddenly deflates
and a longer beep sounds.
6. The measured values
with the WHO blood pressure bar display
beep sounds.
7. If the unit detects an arrhythmia during the measuring process, the
arrhythmia symbol
displayed. You may also hear 8 short beeps; these warn the user of an
irregular heartbeat.
8. If you do not press any buttons for approx. 1 minute, the device
automatically switches to standby mode.
3MAM measuring method
(3 consecutive measurements to calculate an average value)
You have selected the 3MAM measuring method (see section 3, subsection
3.6). You first have to complete any settings you require as per section 3,
subsections 3.3, 3.5 and 3.7.
1. Press the START button
complete, "
" appears on the display. 5 short beeps sound during the
procedure.
automatically inflates until the pressure required for a
. The increasing pressure is displayed
when the device is inflating,
flashes.
,
and
are shown on the display, together
will flash eight times while the reading is
. The device carries out a self test. Once this is

4 Operation

after the device
. At the same time, a longer
GB
27