maintEnancE/clEaning
Optional Heated Humidifier
DeVilbiss Heated Humidifier (cradle and chamber) ...............................................................................................................................................DV5HH
sleePCUbe mainTenanCe
DANGER
Electric shock hazard – Do not attempt to open or remove the enclosure; there are no user-serviceable components
inside. If service is required, contact your home care provider for instructions on obtaining service of your device.
Opening or attempting to service your device will void the warranty.
WARNING
If the operation of the device appears to deviate from normal operation or begins to produce unusual sounds,
discontinue use and contact your home care provider for assistance.
If the device has been dropped, verify functionality of the device. If the device does not function or the enclosure is
damaged, discontinue use and contact your provider.
If the device does not appear to be functioning properly refer to the Troubleshooting section for instructions.
Pressure Accuracy–The SleepCube is designed to provide accurate pressure performance from the factory without additional calibration. No
routine service is required for a period of 2 years of use, provided the device is used in accordance with the manufacturer's directions. After 2
years the device should be inspected.
NOTE–
Some countries require periodic servicing and calibration of this kind of medical device. Contact your home care provider for further information.
Standard Air-Inlet Filter–Check the filter every 10 days and clean it if necessary. Replace the filter every 6 months or sooner if it is damaged.
The standard filter is designed to filter particles larger than 3.0 microns in size. The SleepCube MUST have the standard filter in place during
operation.
Optional Fine Particle Filter–Check the filter every 10 days, and replace it if it is dirty or damaged; otherwise replace the filter every 30 days.
The optional fine particle filter is designed to filter particles as small as 0.3 microns in size and larger.
Filter Maintenance–Refer to Cleaning for instructions.
sleePCUbe CleaninG
WARNING
To avoid electric shock, always unplug the power cord from the wall outlet power source before cleaning.
cAuTION– Never rinse or place the device in water. Never allow liquids to get into or around any of the ports, switches or air filter; doing so will result in
product damage. If this occurs, do not use the device. Contact your home care provider for service.
Device cabinet
1.
Unplug the SleepCube from the power source. Wipe the enclosure with a clean, damp cloth every few days to keep the enclosure dust free.
2.
Allow the device to dry completely before plugging it into a power source.
Filters
Air-Inlet Filter (Standard)
1.
The air-inlet filter should be checked every 10 days and cleaned as necessary. The filter is located in the rear of the device.
cAuTION– Proper filter function is important for the operation of the device and to protect the device from damage.
2.
To clean, remove the dark outer foam filter from the frame.
EN