2. Close the door and turn the latch to the lock position.
Verify the mark on the latch aligns with the lock
symbol and the door is securely closed.
NOTE: Although one power module is usually sufficient
for one operation, Stryker recommends preparing a
second handpiece with power module to ensure quick
intraoperative exchange of handpieces under sterile
conditions.

To Install and Remove a Cutting Tool

1. Push the blade lock button and fully insert the blade
into the blade mount.
2. Verify the FULL INSERT line disappears inside the
blade mount indicating the blade is properly positioned.
www.stryker.com
3. Release the blade lock button.
4. Gently pull on the blade to verify it is secure.
5. If desired, pull and rotate the blade mount to a new
position.
6. Snap the blade mount back into place.
7. Attempt to rotate the blade mount to verify it is secure.
8. To remove the blade, push the blade lock button again.