Operating Instructions Torque Power BL ISO
2
Technical description
Model designation
Transformer primary voltage
Transformer secondary voltage
Cooling air pressure
Cooling air consumption
Spray air pressure
Spray water
Air requirements
Water requirements
Operating conditions
Transport and storage conditions
Protection class
Degree of protection against electric shock
Degree of protection against ingress of water
Mode of operation
Dimensions
Weight (without electric motor and hose)
Weight of transformer
Year of manufacture
62 28 527 D 3547
D 3547.201.01.01.02
Torque Power BL ISO
120 V~ 50/60 Hz
120 V~ 60 Hz
TIYM 24 - 02 M
TDYN - 01 M
100 VA
75 VA
24 VAC
Nominal operating pressure: 44 psi (3 bar),
maximum operating pressure: 73 psi (5 bar)
depending on control pressure
ISO 4-hole connection
Depending on control pressure (min. 18 NL/min,
at nominal operating pressure of 44 psi (3 bar))
2.7 + 0.1 bar (39 psi)
2 + 0.1 bar (29 psi), min. 100 ml/min
Dry, oil-free, dirt-free, not contaminated.
If necessary: Use compressor with dry air system, install
air filter (50 µ), blow through hoses before connnecting
them.
The spray water must be dirt-free according to the appli-
cable regulations. It must not be contaminated.
Ambient temperature: 50 °F – 104 °F (10 °C – 40 °C)
Relative humidity: 30 % – 75 %
Air pressure: 700 hPa – 1060 hPa
Temperature: -40 °F – 158 °F (-40 °C – +70 °C)
Relative humidity: 10 % – 95 %
Air pressure: 500 hPa – 1060 hPa
Protection class I
Type B applied parts
Ordinary equipment (not protected)
Intermittent operation 60 s, 10 min pause
("INT 1/10 MIN")
Width: 150 mm, depth: 130 mm, height: 70 mm
approx. 560 g
approx. 2300 g
OMuu
(on rear side of unit)
2 Technical description