CAUTION
Please separate the transformer from the mains supply
by disconnecting the mains supply plug.
Cleaning and disinfecting the water paths
The Torque Power BL ISO unit and the supply hose for
the BL ISO electric motor are equipped with an internal
water supply which transports the spray from the dental
treatment center to the straight/contra-angle handpiece.
Torque Power BL ISO has no self-cleaning or self-disin-
fecting system for this internal water supply. The internal
spray lines of the Torque Power BL ISO and the supply
hose are, however, connected to the spray supply of the
dental treatment center.
The Torque Power BL ISO spray lines should be
cleaned and disinfected together with the dental treat-
ment center once a week using Dentosept P (Order No.
33 18 156)
as described below:
•
Flush the Torque Power BL ISO with Dentosept P for
at least 20 seconds.
•
Then switch the treatment center and the Torque
Power BL ISO off for 24 hours.
•
Flush the Torque Power BL ISO with water for at
least 30 seconds.
The Torque Power BL ISO and the BL ISO motor must
remain connected to the spray supply of the treatment
center throughout the entire cleaning and disinfection
process.
i
NOTE
Use only instruments that are suitable for the specified
speed range!
CAUTION
Be especially careful during endodontic procedures or
when working near the pulp.
For more information on the cleaning and care of electric
motors, please refer to the operating instructions sup-
plied with the relevant motor.
8 Cleaning and care