3. TURNING ON THE DEVICE

Caution: To avoid
interference, turn on the
Motiva Flora
Port Locator
®
30 cm (12 inches) away from
the Motiva Flora
Tissue
®
Expander product family. A
solid red light will turn on if
you omit this indication.
The device includes an ON
button (Figure 5). A white
light indicates that the Motiva
Flora
Port Locator is on.
®
• Turn on the Port Locator
by pressing the button and
release once the white light
ring is turned ON.
Figure 4: Cleaning and disinfecting the device.
10
Verify out of the
sterile field that the
Motiva Flora
Port
®
Locator's lights come
on when near the
breast expander.
Figure 5: Power button.
OFF
ON
The device will turn
off automatically
after 1 min
without use.