or decreased electromagnetic immunity of this equipment and result in improper
operation.
• The device must not be used in a potentially explosive environment. Do not turn on the
Motiva Flora
Port Locator in potentially explosive environments, such as gas stations.
®
Motiva Flora
Port Locator usage is restricted in areas with fuel, chemical plants, or
®
where blasting operations are in progress.
• DO NOT use the device while it is being charged.
• NEVER use the device when the status light is blinking orange.
ELECTROMAGNETIC SAFETY TESTING
Test
Power input voltage
Electromagnetic radiation
disturbance (radiated
5 VDC
EMISSIONS) CISPR 11.
Motiva Flora
ELECTROSTATIC
DISCHARGE IMMUNITY IEC
to withstand ±8 kV air discharge and ±8
61000-4-2
kV contact discharge without damage.
8
Line frequency
N/A
Port Locator shall be able
®
N/A
EN
Test
Power input voltage
IMMUNITY to proximity
A field of 10 V/m modulated by a 1 kHz
fields from RF wireless
communications equipment
80% AM modulation over the frequency
IEC 61000-4-3 (interim
range of 90 kHz to 80000 kHz.
method)
IMMUNITY to conducted
5V rms, 90 KHz to 80 MHz with an 80%
DISTURBANCES induced
AM with a 1 KHz sine wave AC & DC
by RF fields (conducted RF
power 5V rms sig/control >3 m, 90 KHz
DISTURBANCE IMMUNITY)
to 80 MHz with an 80% AM with a 1 KHz
– a.c. mains IEC 61000-4-6
sine wave.
5V rms, 90 KHz to 80 MHz with an 80%
IMMUNITY to conducted
DISTURBANCES induced
AM with a 1 KHz sine wave AC & DC
by RF fields (conducted
power 5V rms sig/control >3 m, 90 KHz
disturbance immunity) – SIP/
to 80 MHz with an 80% AM with a 1 KHz
SOP PORTS IEC 61000-4-6
sine wave.
Power frequency magnetic
field IMMUNITY IEC 61000-
30 A/m at 60 Hz.
4-8.
Line frequency
N/A
N/A
N/A
N/A