• If the version in the upper right corner has not changed, the SW update failed. Repeat the ADCC SW update procedure. If this update
fails, please contact your local Afinion™ supplier.
• Power must not be disconnected during SW update. Loss of power during SW update might lead to ADCC failure.
• During SW update of ADCC, the unit should be disconnected from the Afinion™ AS100 Analyzer. Tests performed on an Afinion™
AS100 Analyzer connected to an ADCC during SW update might lead to loss of result transfer to HIS/LIS system.

8 ADCC TROUBLESHOOTING

8.1 LED status indicator signals

Symbols
Situations
Power off
Power on
Power on, no HIS/LIS communication. Wait 1 min. Check
ADCC configuration and Ethernet connection
Power on, no Afinion™ AS100 Analyzer communication
Analyzer in Automatic start up procedure. Wait 1 min
after Analyzer has performed Automatic start-up procedure
Afinion™ AS100 Analyzer connection established
HIS/LIS connection established
Connection succeeded, HIS/LIS and Afinion™ AS100
Analyzer connection established
Afinion™ AS100 Analyzer communication failed. Wait
1 min. Analyzer in self-test. Wait 1 min. after Automatic
start-up procedure is performed
HIS/LIS communication fatal error
Unrecoverable error not related to port communication
SW upgrade is finalizing
Factory reset button activated
If the results are not transferred but the LED lights are not displaying any of the errors above, please reboot ADCC. If the error
persists, please contact your local Afinion™ supplier.
For US customers a toll free line for technical support is available
All manuals and user guides at all-guides.com
Power LED
(Green)
OFF
Steady
Steady
Steady
Steady
Steady
Steady
Steady
Steady
Steady
Steady
Single flash
The SW will then be updated on the
ADCC. The update takes about 25
seconds.
The SW update is complete when you
are forwarded automatically to the
configuration page.
The new SW version will be displayed
in the upper right corner.
IOIOI
RS232 LED
Ethernet LED
(Yellow)
(Yellow)
OFF
OFF
Any state
Any state
Any state
OFF
OFF
Any state
Steady
Any state
Any state
Steady
Steady
Steady
Flashing
Any State
Any State
Flashing
OFF
OFF
LED is lit in sequence during upgrade. SW upgrade is success-
fully performed when LED flashes three times simultaneously
Single flash
Single flash
Warning LED
(Red)
OFF
Any state
Any state
Any state
Any state
Any state
OFF
Flashing
Flashing
Flashing
Single flash
7
GB
US