AC Adapter
Input Voltage: AC 100~240V
Input Frequency: 50Hz~60Hz
Output Voltage: DC 5V±5%
Output Current: 2A MAX
SMD chip fuse
32V3A0466003.NR(1206)
Data store and replay
Outline of product
Dimension: 143.3mm(L)X67.4mm(W)X36mm(H)
Weight: ≤140±10g (excluding the batteries)
Box contents:
1. Three AA alkaline batteries
2. One instruction manual
3. One adult finger probe: M-50E012CS09
4. One pediatric finger probe: M-50B008CS09 (optional)
5. One disposable probe: M-50J033CS045 (optional)
6. One USB cable
7. Adapter(optional)
Equipment Classification
Classification according to IEC-60601-1:
According to the type of protection against Electrical shock: Internal electrical power source equipment
and class II equipment;
According to the degree of protection against Electrical shock: Type BF equipment, applied part is
SpO2 probe;
According to the degree of protection against harmful ingress of water: IPX1;
According to the methods of sterilization or disinfection: Non-sterilization, use of liquid surface disin-
fectants only;
According to the mode of operation: Continuous operation.