• Cleaning: Prepare alkaline cleaner solution (e.g.
Neodisher MediClean forte or equivalent) following the
manufacturer's recommendation of approximately 29
mL per approximately 4 L of warm tap water (ratio and
water temperature per manufacturer's recommendation
if applicable). Completely submerge the tubing and elbow
connector in the solution and allow it to soak for a minimum
of ten (10) minutes. At the beginning of the soak time,
flush the inside of the tubing with the prepared solution.
While the tubing is submerged, use a soft bristle brush
(e.g. toothbrush) to brush all accessible areas of the tubing
and elbow connector for a minimum of one (1) minute to
remove any visible signs of debris or dirt.
• Rinse: Rinse the tubing and elbow connector thoroughly
with cool tap water until there is no visible sign of the
alkaline cleaning solution. While rinsing, flush the inside of
the tubing with the cool tap water.
• Visual inspection: Inspect the tubing to ensure that all
debris and dirt have been removed. If there is any sign of
debris or dirt, repeat the steps above until there is no sign of
debris or dirt remaining on or inside the tubing.
• Disinfection: Fully submerge the tubing and elbow
connector in 70% isopropyl alcohol (IPA). Allow it to sit for
a minimum of ten (10) minutes. Flush the IPA through the
tubing, prior to starting the ten (10) minute time.
• Rinse: Rinse the tubing and elbow connector thoroughly
with cool tap water, ensuring that the inside of the tubing is
flushed with water.
• Drying: Dry the tubing and elbow connector outer surface
with a clean low lint towel or cloth and allow the inside of
the tubing to air dry for a minimum of thirty (30) minutes at
room temperature.
PureWick™ Urine Collection System Base with power cord
(Multiple Users)
• Initial Wipe: Wipe the PureWick™ Urine Collection System
base and power cord thoroughly with a clean low lint cloth
dampened with cool tap water. Remove any excess dirt by
wiping the unit with disposable low lint wipes.
• Cleaning: Prepare alkaline cleaner solution (e.g. Neodisher
MediClean forte or equivalent) following the manufacturer's
recommendation of approximately 29 mL
per approximately 4 L of warm tap water (ratio and water
temperature per manufacturer's recommendation if
applicable). Wipe the device and power cord with a low
lint cloth dampened with the solution. Allow the device
and power cord to remain wet for a minimum of ten (10)
minutes, ensuring that the device and power cord are
visibly wet for the ten (10) minutes. Using a soft bristle
brush (e.g. toothbrush), brush all accessible areas of the
PureWick™ Urine Collection System base and power cord for
a minimum of one (1) minute to remove any visible signs of
debris or dirt.
• Rinse: Using a new clean low lint cloth, wipe the PureWick™
Urine Collection S ystem base and power cord thoroughly
with cool tap water to remove the alkaline cleaning solution.
Perform this step until there is no visible sign of the cleaning
solution.
• Visual inspection: Inspect the PureWick™ Urine Collection
System base and power cord to ensure that all debris and
dirt have been removed. If there is any sign of debris or dirt,
repeat the steps above until there is no sign of debris or dirt
remaining on the PureWick™ Urine Collection System base
and power cord.
• Disinfection: Use 70% isopropyl alcohol (IPA) to completely
wet a clean low lint cloth and wipe down the device and
power cord. Allow the disinfecting solution to remain on the
PureWick™ Urine Collection System base and power cord for
a minimum of ten (10) minutes, ensuring the unit is visibly
wet with IPA for ten (10) minutes.
• Rinse: Using a new clean low lint cloth, wipe the PureWick™
Urine Collection System base and power cord thoroughly
with cool tap water to remove the disinfecting solution.
• Drying: Dry the PureWick™ Urine Collection System base
and power cord with a clean low lint towel or cloth.
Replacing Canister and Tubing
Replace the canister and tubing for each patient per facility
protocol or at least every 60 days, whichever is sooner. If you
notice any of the following, replace immediately:
• Canister or tubing has residual urine build-up
• Canister or tubing appear cloudy
• Canister or tubing appear discolored
• Canister becomes cracked
• Tubing becomes torn
• PureWick™ External Catheter no longer securely connects to
collector tubing
Failure to clean and/or replace accessories may affect the
performance of the system.
The useful life of the collection canister and tubing is 60 days.