6 Accuracy table
6 Accuracy table
Volume
range
[μl]
0.5–10
5–50
10–100
20–200
30–300
*A = Accuracy, CV = Coefficient of Variation
20 °C
!
Ex
Final test values based on the nominal volume (= max. volume) printed on
the device and the specified partial volumes at the same temperature (20
°C/68 °F) of the device, surroundings and distilled water, in accordance
with DIN EN ISO 8655.
Operating manual
48
Partial
A* ≤ ±
volume
[μl]
10
1.6
5
2
1
8
50
0.8
25
1.4
5
6
100
0.8
50
1.4
10
4
200
0.8
100
1.4
20
4
300
0.6
150
1.2
30
3
CV* ≤ %
Sub
steps
[μl]
1.0
0.01
2
6
0.4
0.05
0.8
3
0.3
0.1
0.6
2
0.3
0.2
0.6
1.5
0.3
0.5
0.6
1.5
Recommended
tip type
[μl]
0.5–20
2–200
2–200
2–200
5–300
Transferpette® S-8/-12 997208