9 Maintenance
9 Maintenance
9.1 Separate the pipetting unit from the handle
>
When assembling, the pipetting unit must be screwed counter-clock-
wise onto the handle so that it clicks into place.
>
When assembling, do not pull the pipetting unit downwards.
>
Improper handling can lead to damage.
9.2 Disassembly/cleaning
The three main components of the pipetting unit can be easily separated
and disassembled for maintenance, cleaning or replacing parts. The pro-
cedure is explained in detail on the following pages.
Changing the O-rings on the individual shafts is described in detail in the
instructions included with the replacement part.
Operating manual
52
a.
Eject the pipette tips.
b.
To separate, pull the pipetting unit
downward as far as possible. Only
then, turn it clockwise. After one rev-
olution, the unit no longer needs to
be pulled downward while turning.
NOTICE
Transferpette® S-8/-12 997208