2 Safety regulations

2.3 Limitations of use

This instrument is intended for pipetting samples, within the following limitations:
Operating temperature of instrument and reagent should be between +15 °C and +40 °C (59 °F to
104 °F) (other temperatures upon request)
Vapor pressure up to 500 mbar
Viscosity: 260 mPa s 260 mPa s
For viscous media, the speed must be adjusted if necessary.

2.4 Application restrictions

Viscous and wetting liquids may compromise volumetric accuracy. Volumetric accuracy may also be
affected when pipetting liquids whose temperature deviates from the ambient temperature by more
than ± 1 °C/± 1.8 °F.

2.5 Operating exclusions

The user is responsible for checking the compatibility of the device with the intended application. The
device cannot be used:
for liquids that attack polypropylene or polyvinylidene fluoride, polycarbonate/polybutylene
terephthalate, polyether ether ketone, FKM or EPDM (flexible replacement pipette shafts)
Avoid aggressive vapors (risk of corrosion)!
The hand grip is not autoclavable.
2.6 Rechargeable battery and power adapter specifica-
tions
Rechargeable battery
Nickel metal hydride rechargeable battery with 3 cylindrical single cells size AAA, 3.6 V, 700 mAh
Power adapter
Output voltage 6.5 V DC, 200 mA
48
Operating manual
Transferpette® electronic 997209