Standalone Model Operation; Start-Up Procedure - Honeywell E3Point Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Standalone Model Operation

Standalone Model Operation
The E
Point gas monitors (Standalone and Network) are factory configured
3
to conform to specific standards. It is designed to be operational after
physical installation and warm-up procedures.

Start-Up Procedure

Before applying power, check all wiring for continuity, shorts, grounds, etc.
After power-up, ensure the LEDs below are operational as follows:
LED
Description
I/O
Power
1
Built-in sensor
2
Remote Sensor
There is no Alarm C option for the remote sensor (Z2).
14
LCD Screen
Navigation/Programming Buttons
Figure 2. E
Display Modes
Always on = Normal operation
Always off = microcontroller fault or no power
Blinking (twice per second) = self-test
Always on = Alarm A triggered
Slow blink = Alarm B/C triggered (1 blink/second)
Fast blink = Fault (4 blinks per second)
Always off = normal operation
Always on = Alarm A triggered
Slow blink = Alarm B triggered (1 blink/second)
Fast blink = Fault (4 blinks per second)
Always off = normal operation
LED Indicators
Point LCD
3
E
Point Gas Monitor Technical Manual
3

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

1309a00381309a00591309a00511309a0061

Tabla de contenido