Standalone Model Buzzer Menu; Standalone Model Alarm Menu - Honeywell E3Point Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Press enter again to confirm the SetDelay, or use the arrow keys
to select No and cancel the procedure.

Standalone Model Buzzer Menu

The Buzzer menu allows users to select whether or not the built-in audible
alarm can be silenced by pressing a keypad button and which events will
activate the buzzer.
Press enter in the main Buzzer menu screen. The Buzzer screen
displays either Activate or Silence.
Use the arrow keys to scroll to the desired option and press enter
to select.
If Silence is selected, the buzzer can be deactivated (without affecting the
event) by pressing and holding any key on the monitor for 5 seconds.
Select Activate to display available menu options:
B:
The buzzer will activate when alarm level B is reached.
F:
The buzzer will activate upon a system or sensor fault.
Disable:
Buzzer is completely deactivated. However, if there is a
fault alarm, the buzzer will sound one brief beep once every
10 minutes.

Standalone Model Alarm Menu

CAUTION:
!
programming functions of this monitor. Factory settings conform to specific
standards (see Specifications). Any changes made to Alarm Levels may
affect manufacturer's stated standards compliance.
The Alarm menu allows users to define different alarm levels than the
factory settings. Press enter to access the Alarm menu options. If there
is a remote sensor connected, scroll to select either Z1 (built-in sensor) or
Z2 (remote sensor).
Use the arrow keys to scroll to the Alarm A, B, or C option and
E
Point Gas Monitor Technical Manual
3
* Menu *
Activate
Only qualified, knowledgeable personnel should use the
* Menu *
Alarm
Buzzer
Buzzer
Alarm
Z1
Monitor Operation
25

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

1309a00381309a00591309a00511309a0061

Tabla de contenido