Troubleshooting - Honeywell E3Point Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Troubleshooting

If the E
Point Standalone or Network encounters errors, the LCD will
3
display a 2-digit error code. The following table provides an explanation of
these codes. Contact Technical Services if any of these error codes
appear.
Error
Description
Code
01
Main board eeprom error
02
Main board power glitch
03
Sensor board eeprom error
04
Sensor board eeprom uninitialized
05
Sensor board ADC failure
06
Sensor board ADC high saturation/clipping
07
Sensor board ADC low saturation/clipping
08
Sensor span gain not calibrated
09
Sensor not responding to span calibration
10
Sensor unstable while in span calibration
11
Sensor span gain too high
12
Sensor span gain too low
13
Sensor baseline not calibrated
14
Sensor baseline offset too high
15
Sensor baseline offset too low
16
Sensor reflex test (self-test) failed. See
Specifications.
17
Sensor end of life reached
18
Sensor due for calibration
19
Temperature sensor ADC signal too high
20
Temperature sensor ADC signal too low
21
Power overload caused by the presence of two high-powered
sensors
22
Remote sensor communication error
If an O
sensor displays 0% and is in Alarm A status when no nitrogen
2
is present, the sensor may be missing or defective. Contact Technical
Services.
36
Troubleshooting
Sensor Over Range
E
Point Gas Monitor Technical Manual
3
in

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

1309a00381309a00591309a00511309a0061