EN-52
The Filling Configurations are defined below (defaults in Bold).
Configure Item
Auto Tare
Accumulate
Note: please refer to 4.1.3 Accumulation in Weighing section for how to perform accumulation.
4.5.3 Input/Output (I/O) Setup
The I/O's setup can be customized for various user preferences.
The I/O's setup is defined below (defaults in Bold).
Item
Enable
Discrete Input 1
Discrete Input 2
Discrete Output 1
Discrete Output 2
Discrete Output 3
Discrete Output 4
Note: The I/O's setup will only work when the I/O Option board has been installed. See the Accessory
list in section 9.2 for information. The option I/O board provides two isolated inputs and four dry-contact
normally open relay outputs which can be used for simple process weighing.
Press the button
Menu.
With the button corresponding to the icon
go down the list and highlight Application
Mode. Enter this sub-menu by pressing the
Softkey corresponding to the icon
Find Quality Products Online at:
Option(Bold is default)
Off/On
Off/Manual/Automatic
Available Settings
On, Off
Off, Zero, Tare, Clear Tare, Print, Start/Stop, Pause/Continue
Off, Zero, Tare, Clear Tare, Print, Start/Stop, Pause/Continue
Off, SP1, SP2, SP3, SP4, Alarm, Zero
Off, SP1, SP2, SP3, SP4, Alarm, Zero
Off, SP1, SP2, SP3, SP4, Alarm, Zero
Off, SP1, SP2, SP3, SP4, Alarm, Zero
to enter the Main
.
GlobalTestSupply
www.
Description
Off: Auto tare is turned off.
On: The first stable weight (>=5d) will be tared as a
container automatically.
Off: The Softkey "∑" doesn't display.
Manual: The Softkey "∑" displays. User can press
the key to accumulate the stable weight.
Automatic: The Softkey "∑" displays. The weight
will be accumulated automatically.
Note:
The load to be accumulated has to be >= 5d and
the next accumulation can only start once the pan
has been cleared.
When LFT is ON (no such limit when LFT is OFF
or the approved model is OIML),
a. gross weight and net weight cannot be
accumulated;
b. gross 0 has to be reached before a new sample
accumulation.
When LFT is OFF,
a. gross weight and net weight can be accumulated
at the same time;
b. a new sample can be accumulated after
reaching gross 0 or net 0.
,