5. LCD / Time Settings:

5.1
LCD contrast
(7)
Can be adjusted between 1 and 9.
5.2
Backlight brightness
(7)
Can be adjusted between 1 and 9
5.3
Date
(Optional)
First two digits:
day
Middle two digits:
month
Last two digits:
year
5.4
Time
(Optional)
First two digits:
hour
Last two digits:
minute

6. Power management settings:

6.1 Empty weighing plate special display
(On)
If it is ON: there will be a flashing point on display
instead of weight value if scale is powered by
battery and empty during 30 seconds. (in a scale
with LCD display "
" is shown)
------
If it is OFF: does not work.
6.2 Power management
(On)
Manages empty weighing plate special display
mode and the other power saving modes.
If it is ON: works for saving power in all modes.
If
it
is
OFF:
works
without
any
power
management in all modes.

7. Filters:

7.1 Median filter
(On)
7.2 Average filter
(On)
7.3 Point position
It shows the location of point on the unit weight
display. If the number is " 0 " point is at the end of
the right side. If it is " 5 " point is at the end of the
left side.

8. Calibration:

8.1 Calibration coefficient
(It can be changed, if the calibration key is ON)
8.2 Maximum capacity
(It can be changed, if the calibration key is ON)
8.3 Division
0: 3000
1: 2 x 3000
(division can not be set if it is approved)
3: 6000 / 7500 (division can not be set if it is
approved)
4: 12000 / 15000(division can not be
set it is approved)
6: 1500
(It can be changed, if the calibration key is ON.)
8.4 Calibration empty weighing plate internal
count
(It can be changed, if the calibration key is ON.)
8.5 Maximum capacity unit
0: g
1: kg
2: ton
(It can not be changed, if it is approved)
8.6 Gravity value of place where instruments
was calibrated
(It can be changed, if the calibration key is ON)
8.7 Gravity value of place where instrument
will be used
(It can be changed, if the calibration key is ON)
8.8 Load Cell mV/ V value

9. Relay settings:

Not in this type.

10. Interface settings:

(Only for PRINTER/PC type scales.)
10.1
Communication ON/OFF (On)
If it is ON :RS232 communication is active.
If it is OFF : RS232 communication is inactive.
(It must be activated in only for printer/pc type
scales, otherwise
ER 8
message will be seen.)
10.2 ID Number
Not in this type.
10.3 Continuous transmitting
Not in this type.
10.4 PC data package selection
Not in this type.
10.5
Baud Rate
It is communication speed between Computer
and Scale
(1)
0: 9600
1: 19200
2: 38400
3: 57600
4: 115200
- 11 -

11. Testing Menu:

Not in this type.

12. Others:

12.1 Approved
(Off)
(It can be changed, if the calibration key is ON.)
If it is turned ON, all settings depended on
(1)
ERROR1 : Turn off the scale and then turn on it again. If failure continues, call service.
ERROR2 : Wait for a while. Turn off the scale and then turn on it again. If failure still continues, call
service.
(9.8022)
ERROR3 : Wait for a while. Sure weighing plate is empty and press
off the scale and then turn on it again. If failure still continues, call service.
ERROR4 : Turn off the scale and then turn on it again. If failure continues, call service.
(9.8022)
ERROR5
Be sure the weighing plate is empty. (If the weighing plate is not on scale, put it on to the scale)Turn off
the scale and turn it on again, when the weighing plate is empty.
(2,0)
Call the service if this warning is active again, when the weighing plate is empty.
ERROR6 : Turn off the scale and turn on it again. If failure continues, call service.
ERROR7 : Turn off the scale and then turn on it again. If failure continues, call service.
ERROR8 : Interface communication error. Turn off the scale and then turn on it again. If scale is not
pc/printer type and error still continues, call the service.
If scale is pc/printer type, change SET 10.1 to "On" and turn off the scale and turn on it again. If still
continues call the service.
ERROR9 : Turn off the scale and then turn on it again. If failure continues, call service.
ERROR11
ERROR12
ERROR13
( )
There will be no data transmission to the PC/Printer and new transaction will be blocked until alibi
memory erased.
Erase alibi memory, turn off the scale and then turn on it again. For Printer type models see related
( )
documents.
ERROR14
connector is unplugged, plug it again. If failure continues, turn off the scale and then turn on it again. If
failure still continues, call service.
( )
ERROR15
failure still continues, call service.
( )
ERROR16
Lo bAtt: The battery is empty. Scale does not work. It should be charged or changed as soon as
possible.
Full (on weight display): Weight is more than maximum capacity.
Full (Price display): Price to pay is too long to shown on display.
nnnnnn : Platform (load carrier) is not fitted, or when there is weight on the platform weighing
instrument may be tried to open up. On the empty platform. (The platform is not installed, replace) If
necessary, there was no load on the instrument a more open platform.
If load carrier has two platform, you make sure to empty on load carrier.
condition "APPROVED" returns to factory
settings values.
12.2 Starting delay (10sn)
(On)
(If it is approved, it cannot be changed)

THINGS TO DO BEFORE CALLING THE SERVICE

FAILURE SIGNALS
: Initial zero setting range control error. Weighing not allowed.
: Turn off the scale and then turn on it again. If failure continues, call service.
: Turn off the scale and then turn on it again. If failure continues, call service.
: For pc/printer type models, If alibi memory is full, this message will be shown.
: Loadcell cable detection error. Check loadcell connector and cable. If the
: Wait for a while. If failure continues, turn off the scale and then turn on it again. If
: See "PrinterUserManualAnnex.doc". If failure continues, call service.
- 12 -
O
key. If failure continues, turn