Descargar Imprimir esta página

Packaging; Additional Information - Covidien TipVision Instrucciones De Uso

Publicidad

Idiomas disponibles

Idiomas disponibles

PACKAGING

InstruSafe Systems may be wrapped with a legally marketed wrap or placed in a legally marketed rigid container.
Refer to the Indications for Use statements found in the instructions for use.
Using with Wrap:
1. Before wrapping the cassette, ensure it has the proper foot style. R should be used with wrap. See Figure F.
Figure F.
'R' or Rounders Corner Guards (for wrap)
'TF' or Tall Foot (for Rigid Containers)
2. Wrap the cassette using legally marketed wrap per the wrap manufacturer's instructions.
3. Sterilize the pack using one of the sterilization cycles listed in the instructions for use.
Using with Rigid Container:
Refer to the Indications for Use statements found in the instructions for use.
1. Before placing the cassette in a rigid container ensure it has the proper foot style. TF should be used with a rigid container.
See Figure F.
2. Place the cassette into the rigid container. Follow the container manufacturer's instructions for sealing the container.
Note: Do not wrap the cassettes before placing into the container for sterilization.
3. Sterilize the container and contents using one of the sterilization cycles listed in the instructions for use.

Additional Information:

To remove the cassette from a rigid container aseptically, follow container manufacturer's instructions for use.
1. With gloved hands, place palms outward and reach through the openings in the cassette cover.
2. Grasp the cassette handles and lift to their raised position. See Figure G.
Figure G.
3. Tilt the handles inward, away from the sides of the rigid container, and lift the cassette out of the container being careful to
not touch the top or outside of the container.
4. Place cassette on a sterile surface.
6

Publicidad

loading