5.7 Measurement of resistance

WARNING
Do not perform measurements on the circuit under the
voltage. Before the measurement disconnect the pow-
er and discharge capacitors.
To perform measurement of resistance:
 set the rotary switch at Ω
 connect black test lead to COM terminal, and red test lead to
VΩLoZV
 contact the tips of test probes to the points of measurement; the
best solution is to disconnect one side of the tested element, to
prevent the remaining part of the circuit interfere with the read-out
of the resistance value,
 read the measurement result on the display.

5.8 Circuit continuity test

WARNING
Do not perform measurements on the circuit under the
voltage. Before the measurement disconnect the pow-
er and discharge capacitors.
To perform the continuity test:
 set the rotary switch at Ω
 connect black test lead to COM terminal, and red test lead to
VΩLoZV
 press MODE button, to display Ω
 contact the tips of test probes to the points of measurement,
 read the measurement result on the display; the beep will be acti-
vated when resistance values are below approx. 50 Ω.