Descargar Imprimir esta página

Sonel CMP-1015-PV Manual De Instrucciones página 72

Publicidad

Idiomas disponibles

Idiomas disponibles

DC voltage measurement
Range
600.0 mV
6.000 V
60.00 V
600.0 V
1500 V
Input impedance: 10 MΩ
Overload protection: 1500 V DC
AC+DC voltage measurement
Range
6.000 V
60.00 V
600.0 V
1000 V
Input impedance: 10 MΩ
Frequency range: 50 Hz...1000 Hz
Overload protection: 1000 V DC/AC RMS
LoZ measurement
Range
6.000 V
60.00 V
300.0 V
600.0 V
1000 V
All AC voltage ranges are specified from 10% to 100% of range
Input impedance: 300 kΩ
Overload protection: 1000 V DC/AC RMS
70
Resolution
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
Resolution
0.001 V
0.01 V
0.1 V
1 V
Resolution
0.001 V
0.01 V
0.1 V
0.1 V
1 V
CMP-1015-PV – USER MANUAL
Accuracy
 (0.8% m.v. + 8 digits)
 (0.5% m.v. + 5 digits)
 (0.8% m.v. + 5 digits)
Accuracy
 (1.5% m.v. + 20 digits)
 (2.5% m.v. + 20 digits)
Accuracy
 (3.0% m.v. + 40 digits)
unspecified

Publicidad

loading

Este manual también es adecuado para:

Wmgbcmp1015pv