"m.v." means a standard measured value.
True RMS measurement for AC current
Range
60.00 A
600.0 A
1000 A
All AC current ranges are specified from 10% to 100% of range
Frequency range: 50 Hz...60 Hz
Overload protection: 1000 V DC/AC RMS
DC current measurement
Range
60.00 A
600.0 A
1000 A
Overload protection: 1000 V DC/AC RMS
True RMS voltage measurement
Range
6.000 V
60.00 V
600.0 V
1000 V
All AC voltage ranges are specified from 10% to 100% of range
Input impedance: 9 MΩ
PEAK function accuracy: 10% m.v.
Frequency range: 50 Hz...1000 Hz
Overload protection: 1000 V DC/AC RMS
11 Technical data
Resolution
0.01 A
0.1 A
1 A
Resolution
0.01 A
0.1 A
1 A
Resolution
0.001 V
0.01 V
0.1 V
1 V
CMP-1015-PV – USER MANUAL
Accuracy
(2.5% m.v. + 5 digits)
Accuracy
(2.0% m.v. + 8 digits)
Accuracy
f = 50...60 Hz
(1.2% m.v. + 5 digits)
f = 61...1000 Hz
(2.5% m.v. + 5 digits)
69