Technical information is subject to change without notifica-
tion to allow for updates.
• This device complies with European Standard EN60601-
1-2 (In accordance with CISPR 11, IEC61000-3-2,
IEC61000-3-3, IEC61000-4-2, IEC61000-4-3, IEC61000-
4-4, IEC61000-4-5, IEC61000-4-6, IEC61000-4-7,
IEC61000-4-8, IEC61000-4-11, EN55024) and is subject
to particular precautions with regard to electromagnetic
compatibility. Please note that portable and mobile HF
communication systems may interfere with this unit.
• The device complies with the EU Medical Devices
Directive 93/42/EEC, the German Medical Devices Act
(Medizinproduktgesetz) and the standards EN1060-1
(Non-invasive sphygmomanometers – Part 1: General
requirements), EN1060-3 (Non-invasive sphygmoma-
nometers – Part 3: Supplementary requirements for elec-
tro-mechanical blood pressure measuring systems) and
IEC80601-2-30 (Medical electrical equipment – Part 2
– 30: Particular requirements for the basic safety and
essential performance of automated non-invasive sphyg-
momanometers).
• The accuracy of this blood pressure monitor has been
carefully checked and developed with regard to a long
useful life. If using the device for commercial medical pur-
poses, it must be regularly tested for accuracy by appro-
priate means. Precise instructions for checking accuracy
may be requested from the service address.

11. Mains adapter

Model no.
LXCP12-006060BEH
Input
100 – 240 V, 50 – 60 Hz, 0.5 A max
Output
6 V DC, 600 mA, only in connection with beurer
blood pressure monitor.
Supplier
Shenzhen Iongxc power supply co., ltd
Protection This device is double insulated and protected
against short circuit and overload by a primary
thermal fuse.
Make sure to take the batteries out of the com-
partment before using the mains adapter.
Polarity of the the DC voltage
connection
Double insulated / equipment class 2
Enclo-
Equipment enclosed to protect against con-
sures and
tact with live parts, and with parts which can
Protective
become live (finger, pin, hook test).
Covers
The operator shall not contact the patient and
the output plug of AC mains adapter simulta-
neously.